Firmware Upgrade
11
8. Click Connect.
In the Atmel Flip” window, in the Operations Flow column, the Run
button is active, and the name of the chip appears as the name of the third
column: AT89C51RD2.
Verify that in the Buffer Information column, theHEX File: PL8.hex
appears.
Figure 12: Atmel Flip Window (Connected)
9. Click Run.
After each stage of the operation is completed, the check-box for that
stage becomes colored green
1
.
When the operation is completed, all 4 check-boxes will be colored green
and the status bar message: Memory Verify Pass appears
2
1 See also the blue progress indicator on the status bar
:
2 If an error message: “Not Finished” shows, click Run again